1. The biochemical incubator must be placed in a sun-proof, ventilated and dry place, with the distance between the equipment and the wall not less than 10cm.
2. The biochemical incubator should be installed on a three-pin power socket with reliable grounding to ensure normal operation and personal safety.
3. In order to maintain the beauty of the equipment, it is not allowed to use acid, alkali and other corrosive items to scrub the inner and outer surfaces of the box. It can be cleaned regularly with a neutral detergent and wiped with a dry cloth.
4. When replacing the fuse, the same fuse should be replaced according to the fuse current and size requirements in the fuse mark.
5. When the biochemical box is not in use, the power should be turned off and the box should be kept clean and dry.
6. The dust and dirt on the back condenser should be cleaned every six months or so to ensure good heat dissipation conditions during the operation of the machine.