Features:
[Vacuum speed] 15 minutes to pump to -60kpa
[Air Replacement] Automatic air exchange, 5 air exchange cycles, 15 minutes apart--times
[Fully automatic control system] One-button start to end, the whole process is automatically completed
[Display status] Programmable controller, touch screen display temperature, humidity, pressure, sterilization time, residual gas processing time and other data
[Printer] uses a micro printer to print out various data every minute during the entire sterilization process,
[Humidity sensor] The built-in humidity sensor automatically detects the humidity in the cavity to ensure that the humidity in the cavity is not less than 45%rh
[Humidification system] Pulse humidification under vacuum to ensure uniform and controllable humidity in the cavity; humidity in the disinfection room is controlled within the range of 55% ± 20% relative humidity;
[Sealed door] 200Kpa pressure resistance, no leakage of EO gas.
[Residual gas treatment device] The filtration rate of the filter for particle diameter ≥ 0.3μm is not less than 99.5%.
[Vacuum rate] Under no-load conditions, the time for the sterilizer to vacuum from atmospheric pressure to -60ka is 30 minutes
[Automatic dosing] Sterilization chamber built-in piercing technology to ensure sufficient gas, safe and reliable
[Biological indicator incubator] A special constant temperature incubator for biological indicators controlled by a microcomputer is used for the constant temperature incubation process of biological indicators for sterilization effect monitoring. The temperature control accuracy is ≤+0.3°C within the full range; the temperature can be self-controlled Set up.
[Heating system] Use explosion-proof electric heating device
[Multi-protection function] With over-voltage and over-temperature protection function, it is safer and more reliable
[Alarm system] There are high temperature alarm system, ultra-low pressure alarm system
【Vacuum system】Using oil-free vacuum pump, simple and convenient, fast vacuum rate, long service life.
[Disinfection room leakage rate] The leakage rate is measured under the conditions of no-load, the sterilization room temperature is unchanged, the test pressure is -50kPa, and the measurement time is 1h. The average leakage rate during the measurement time is ≤0.16kPa/min.
Applicable range:
Medical industry: Hysteroscopy, laparoscopy, laryngoscope, probe, rigid endoscopy
Speculum, hose endoscope, ophthalmic lens, fiber optic cable,
Intracranial pressure sensor, cold therapy probe, prostatectomy device, esophageal dilator, electrocautery equipment,Defibrillation electrode, laser handpiece
Metal, glass, silica gel, rubber, Chinese medicine and western medicine, human implants, DNA, etc.
Specifications:
【Product Model】MQ-H120, MQ-H190,
【Product form】Manual door vertical
【Heating method】Electric heating
【Power rate】2.8KW.3.8KW
【Power source】AC220V, 50Hz
【Effective volume】120L/190L
【Weight】170kg/260kg
【Temperature display】LCD touch screen
【Disinfection room material】 304 stainless steel
【Select gas】 Pure ethylene oxide gas
【Temperature during disinfection】50℃ ±3℃
【Working pressure】-60kpa
【Disinfection anti-pressure value】 -140kpa
【Humidity during disinfection】45%-75%
【Disinfection time】 0-99 hours adjustable
Features:
[Vacuum speed] 15 minutes to pump to -60kpa
[Air Replacement] Automatic air exchange, 5 air exchange cycles, 15 minutes apart--times
[Fully automatic control system] One-button start to end, the whole process is automatically completed
[Display status] Programmable controller, touch screen display temperature, humidity, pressure, sterilization time, residual gas processing time and other data
[Printer] uses a micro printer to print out various data every minute during the entire sterilization process,
[Humidity sensor] The built-in humidity sensor automatically detects the humidity in the cavity to ensure that the humidity in the cavity is not less than 45%rh
[Humidification system] Pulse humidification under vacuum to ensure uniform and controllable humidity in the cavity; humidity in the disinfection room is controlled within the range of 55% ± 20% relative humidity;
[Sealed door] 200Kpa pressure resistance, no leakage of EO gas.
[Residual gas treatment device] The filtration rate of the filter for particle diameter ≥ 0.3μm is not less than 99.5%.
[Vacuum rate] Under no-load conditions, the time for the sterilizer to vacuum from atmospheric pressure to -60ka is 30 minutes
[Automatic dosing] Sterilization chamber built-in piercing technology to ensure sufficient gas, safe and reliable
[Biological indicator incubator] A special constant temperature incubator for biological indicators controlled by a microcomputer is used for the constant temperature incubation process of biological indicators for sterilization effect monitoring. The temperature control accuracy is ≤+0.3°C within the full range; the temperature can be self-controlled Set up.
[Heating system] Use explosion-proof electric heating device
[Multi-protection function] With over-voltage and over-temperature protection function, it is safer and more reliable
[Alarm system] There are high temperature alarm system, ultra-low pressure alarm system
【Vacuum system】Using oil-free vacuum pump, simple and convenient, fast vacuum rate, long service life.
[Disinfection room leakage rate] The leakage rate is measured under the conditions of no-load, the sterilization room temperature is unchanged, the test pressure is -50kPa, and the measurement time is 1h. The average leakage rate during the measurement time is ≤0.16kPa/min.
Applicable range:
Medical industry: Hysteroscopy, laparoscopy, laryngoscope, probe, rigid endoscopy
Speculum, hose endoscope, ophthalmic lens, fiber optic cable,
Intracranial pressure sensor, cold therapy probe, prostatectomy device, esophageal dilator, electrocautery equipment,Defibrillation electrode, laser handpiece
Metal, glass, silica gel, rubber, Chinese medicine and western medicine, human implants, DNA, etc.
Specifications:
【Product Model】MQ-H120, MQ-H190,
【Product form】Manual door vertical
【Heating method】Electric heating
【Power rate】2.8KW.3.8KW
【Power source】AC220V, 50Hz
【Effective volume】120L/190L
【Weight】170kg/260kg
【Temperature display】LCD touch screen
【Disinfection room material】 304 stainless steel
【Select gas】 Pure ethylene oxide gas
【Temperature during disinfection】50℃ ±3℃
【Working pressure】-60kpa
【Disinfection anti-pressure value】 -140kpa
【Humidity during disinfection】45%-75%
【Disinfection time】 0-99 hours adjustable